Patients › Shoulder

Suture Anchors and How Repairs Are Held to Bone

19 citations

What is it

A suture anchor is a small device that holds stitches into bone. It solves a problem that stitches alone cannot: you can sew soft tissue to soft tissue, but you cannot sew soft tissue to bone. The anchor provides something for the stitches to hold on to.

Almost every repair in the shoulder uses them. When a rotator cuff tendon is reattached, when a torn labrum is put back onto the rim of the socket, when a ligament is reattached after a dislocation — in each case an anchor is placed into the bone and its stitches are passed through the tissue and tied down.

The anchor is not what holds the repair together in the long run. It holds the tissue still against the bone for the weeks it takes the two to knit together biologically. Once healing has happened, the join is living tissue, not hardware.

What it's made of

Anchors have been made from several different materials over the years, and the field has moved through them in a fairly clear sequence.

The earliest were metal — usually titanium. They held extremely well but stayed in the bone permanently and showed up on every X-ray afterwards, which made later imaging harder to interpret.

Bioabsorbable anchors followed, made from polymers designed to dissolve over a few years. The intention was sound, but some of the early materials broke down unpredictably and could leave a cavity in the bone or provoke inflammation.

PEEK — a strong medical-grade plastic — is widely used now. It does not dissolve, but it does not show up on X-ray and its stiffness is closer to bone than metal is.

All-suture anchors are the newest, and are now in widespread use in the shoulder. There is no rigid body at all: a small sleeve of suture material is passed into a narrow hole and then tightened so that it bunches up and grips inside the bone. The hole needed is considerably smaller than for a solid anchor, so much less bone is removed.

The stitches themselves have changed as much as the anchors. Modern sutures are braided from an extremely strong polyethylene fibre, and are now stronger than the tendon they hold.

How your surgeon uses it

Anchors are used in several quite different shoulder operations, and what the surgeon is trying to achieve is not the same in each.

Rotator cuff repair

Here the job is to hold a broad, flat tendon down onto an equally broad area of bone — the footprint — so the two can knit together across that whole surface.

There is more than one reasonable way to arrange this, and practice differs between surgeons. A single-row repair places one line of anchors along the edge of the bone where the tendon should sit. A double-row or suture-bridge construct adds a second row further down the bone, so the stitches from the inner row are carried outwards and tensioned over the tendon, pressing a broad surface flat against the bone rather than holding it at a few points.

The second arrangement is thought to favour healing, and it does produce fewer retears on follow-up scans — but as the evidence below sets out, that advantage shows up more reliably on the scan than in how shoulders actually feel. Which is used depends on the size and shape of the tear, the quality of the bone and tendon, and the judgement and training of the individual surgeon.

Labral repair

The labrum is a rim of cartilage around the edge of the shoulder socket. It is not being held onto a flat surface but reattached to the edge of one, so the anchors are smaller and are placed around the rim of the socket, usually three or four of them, with the stitches passed around the labrum to draw it back against the bone.

This is the repair done for shoulder instability after a dislocation, where the labrum has been torn off the front of the socket. The socket is small and the bone is thin, so anchor size and how much bone each one takes up matter more here than almost anywhere else in the shoulder — which is part of why all-suture anchors have been taken up readily for this operation.

Biceps tenodesis (loop and tack)

Sometimes the long head of biceps tendon is detached from inside the joint and reattached to the arm bone a little lower down. The requirement is different again: not spreading a tendon over a surface, and not restoring a rim, but holding one cord-like tendon securely at a single point while it heals into the bone.

One straightforward way to do this is often called loop and tack. A stitch is passed around the tendon and locked onto itself, so it grips the tendon like a noose rather than relying on the thread pulling through tendon fibres. That loop is then tacked down to the bone with a single small anchor.

The alternative approach wedges the tendon into a drilled socket with a screw. Loop and tack avoids drilling a socket of that size, so less bone is removed, and the grip on the tendon comes from the looped stitch rather than from compression against a screw.

A note on knots

In any of these repairs, the stitches may be tied in knots, or held by a locking mechanism built into the anchor so that no knot is needed. Both are in routine use, and the evidence does not favour either.

What to expect

The anchors stay in. They are not removed once healing is complete, and they do not need to be. Removing them would mean another operation and would take bone with them.

They will not set off airport scanners. All-suture and PEEK anchors contain no metal at all. Even titanium anchors are far too small to trigger a walk-through detector.

You can have an MRI. Anchors in current use are MRI-safe. Metal anchors can cause a local blur on the image near the anchor, which is one of the practical advantages of the newer materials — the scan is easier to read if your shoulder ever needs imaging again.

They are not what limits your recovery. The anchor is secure from the moment it is placed; it is the tendon-to-bone healing that takes months, and that is what your sling and rehabilitation timeline are protecting. Recovery follows the timeline of the operation itself, so see the page for your specific procedure.

Occasionally an anchor can loosen or pull out, most often in soft bone. This is uncommon, and it is one of the reasons the choice of anchor is matched to the quality of your bone.

What the evidence says

The honest summary is that anchors work well and that the differences between the various types and techniques are smaller than the amount written about them would suggest.

All-suture anchors hold as well as solid anchors in testing and in patients, while removing less bone — which is the main reason for the shift towards them. Comparisons of the newer suture tapes against conventional round sutures have not shown a clear clinical advantage, despite a sound theoretical case for them. Tying knots and using knotless designs give much the same result.

Double-row repairs do appear to improve the chance of the tendon healing to bone on a follow-up scan, but that has not translated into a clear difference in how shoulders feel and function. That gap between what the scan shows and what the patient reports is a recurring theme in shoulder surgery.

The materials story is the one place where the evidence is genuinely one-sided: the early bioabsorbable anchors produced meaningfully more problems than what replaced them, which is why they are largely no longer used.

If you want the detail — how anchors developed, what the sutures are actually made of, and how all this compares with the older technique of passing stitches through tunnels drilled in the bone — that is in the section below.

Advanced reading: the deeper science (optional)

This section goes further than you need for your own treatment decisions. Anchors are worth the extra reading because they are the clearest example in shoulder surgery of engineering outrunning biology: over four decades the fixation has become far stronger, and the thing that fails has moved each time — from the stitch, to the anchor, to the tendon itself.

The problem anchors solve

Sewing tendon to tendon is straightforward. Sewing tendon to bone is not, because bone will not hold a stitch — the thread simply cuts through the surface.

The original solution was transosseous repair: drill tunnels through the bone and pass the sutures through them, so the thread is anchored around a bridge of solid bone rather than gripping its surface. Done open, this was the standard for decades and it worked.

Arthroscopy is what displaced it. Through keyhole portals you cannot reliably drill and thread a tunnel across a piece of bone you can only see from one side. The anchor exists because it lets a surgeon achieve fixation from a single direction, through a portal — not because it was shown to be better than tunnels.

That distinction matters, because the comparison has since been done. Biomechanically, anchor fixation produces less gap formation under cyclic loading than transosseous repair, with no significant difference in ultimate load to failure [1]. And now that arthroscopic techniques have been developed to recreate bone tunnels without an implant, the clinical comparison is available too: arthroscopic transosseous repair gives comparable clinical and structural outcomes to anchor repair, its only measured advantage being slightly better recovery of abduction [2].

So the tunnel was never really beaten. It was made impractical by the shift to keyhole surgery, and it is now partly returning.

Four generations of anchor

Metal. The first anchors were titanium or stainless steel, and they held superbly. Their drawbacks were downstream: they stay in the bone permanently, and they scatter on later imaging, which matters in a joint likely to be scanned again.

Bioabsorbable. Polymer anchors designed to dissolve over a few years followed, and the shift was rapid and near-total — a 2007 review describes a major shift from metallic to bioabsorbable anchors, attributing it to higher complication rates with metal [3].

That verdict did not entirely survive contact with the polymers themselves. The early poly-L-lactic-acid materials degraded slowly and unpredictably, and could provoke inflammation, leave a fluid-filled cavity in the bone, or shed fragments into the joint. The more measured assessment a few years later was that bioabsorbable anchors remain safe and reproducible, with complications amounting to a fraction of the total implanted — while emphasising that meticulous insertion technique is what keeps them that way [4]. Both reviews stress the same specific point: the anchor must be sunk below the level of the cortex, because a proud anchor abrades whatever passes over it [3].

PEEK. A high-performance thermoplastic that is radiolucent, non-degrading, and closer to bone in stiffness than metal is. It is now the standard solid-anchor material, and is worth a section of its own below.

All-suture. The current generation abandons the rigid body entirely. A sleeve of suture material is passed through a small drill hole and then tensioned so that it deforms and locks against the inner wall of the bone.

Across the design generations the raw numbers improved sharply — newer anchors showed markedly increased load-to-failure strengths — with the important caveat that an anchor which fails by pulling out at low load risks becoming a loose body inside the joint [5].

What PEEK actually is

Polyether ether ketone is a semi-crystalline thermoplastic from the polyaryletherketone family. It is not a plastic in the everyday sense: it is used in aerospace and in spinal implants, it melts around 340°C, it tolerates repeated steam sterilisation, and it is chemically inert in the body.

Three properties explain why it displaced both metal and the early absorbables as a solid-anchor material.

It is radiolucent. PEEK does not show on X-ray and produces no scatter on CT or MRI. A shoulder repaired with PEEK anchors can be imaged afterwards and read cleanly — a genuine advantage in a joint that may well be scanned again.

Its stiffness is close to bone. PEEK's elastic modulus is in the region of cortical bone, where titanium's is roughly ten times higher. A very stiff implant in less stiff bone concentrates stress at the interface; matching the modulus more closely spreads it.

It does not degrade. Unlike the absorbable polymers it replaced, PEEK does not break down, so there is no resorption phase during which the anchor weakens or the body reacts to its breakdown products.

What PEEK does to bone, which is not nothing

The natural assumption is that an inert, non-degrading material provokes no biological response. The imaging evidence complicates that.

Perianchor cysts — fluid-filled cavities in the bone immediately around the anchor — occur with every anchor type. Comparing all-suture, bioabsorbable and PEEK anchors across 213 repairs, cysts formed in 10.8% of cases overall, with no significant differences between the three materials in clinical scores, retear rates or perianchor bone reaction [6].

A more granular comparison is less flattering to PEEK. Across 73 repairs, despite both being non-absorbable, all-suture anchors produced less osseous reaction than PEEK — with the authors concluding that anchor choice should weigh not only initial fixation strength but the post-operative biological response [7].

The reassuring part is what those cysts appear to mean. Around PEEK anchors specifically, cyst formation stabilises by six months and shows no association with functional scores or with retear [8]. So this is a radiographic finding rather than a clinical problem — which matters mainly because it will be reported on your scan and can look alarming.

Where it does have consequences is revision. Anchorless transosseous repair, which leaves no implant at all, avoids peri-implant cyst formation entirely — and the authors argue the real advantage is in the prospect of additional surgery, where undisturbed bone gives more options [9]. That is the same bone-stock argument that drives the move to all-suture anchors, arrived at from a different direction.

Why all-suture anchors won on bone, not on strength

The intuition is that a soft anchor must be weaker than a solid one. It is not. All-suture anchors have similar or better mechanical properties than regular anchors, with a low-profile design that preserves bone tissue, and clinical series show satisfactory results with low complication rates [10].

The real argument for them is the hole. A solid anchor needs a socket sized to its body; an all-suture anchor needs a hole roughly half the diameter. In the greater tuberosity — where the bone is often soft, and where a revision may one day need somewhere fresh to place fixation — how much bone was removed the first time is a genuine consideration.

The clinical evidence supports using them where it matters most. Double-row suture-bridge cuff repair using all-suture anchors for the medial row produced similar excellent outcomes to the same repair using solid medial-row anchors [11].

They are not free of the cyst phenomenon either — around 40% of patients after double-row suture-bridge repair with all-suture anchors, with a low anchor insertion angle and a large mediolateral tear identified as risk factors [12]. That insertion-angle finding is a technical point with a practical edge: how the anchor is aimed is something the surgeon controls.

The suture became the strongest part, and that moved the problem

The less visible revolution is in the thread. Modern sutures are braided from ultra-high-molecular- weight polyethylene — the FiberWire class of materials — and the result is a suture substantially stronger than the tissue it passes through.

This solved one problem and created another. Repairs no longer fail by the suture snapping. They fail by the suture cutting through the tendon, like cheese wire — which is why so much design effort has gone into spreading the load rather than increasing strength further.

Flat suture tape is the direct response: a wider footprint against the tendon means lower pressure per unit area for the same tension. The theory is sound and the laboratory agrees. The patients do not, quite: although biomechanically superior, suture tapes showed similar retear rates and postoperative function to conventional round sutures [13].

That is worth sitting with, because it is the same pattern as the anchor comparisons. Once fixation is strong enough, making it stronger stops changing the outcome — the limiting factor has become the biology of tendon healing to bone, which no implant addresses.

Configuration: where the remaining differences are

Single-row repair places one line of anchors at the tendon edge. Double-row and suture-bridge constructs add a second, more lateral row, compressing a broad area of tendon against bone. Across the evidence, double-row and suture-bridge repairs have lower retear rates than single-row repair in most tear-size categories, with no difference between double-row and suture bridge [14].

Knot-tying versus knotless designs show no significant difference in retear rates [15].

Two ways a repair fails, and only one of them is fixable

This is the most consequential thing in this section, and it is rarely explained to patients. A cuff repair that fails does not simply "fail" — it fails in one of two patterns, and they have very different implications.

Type 1 is failure at the footprint. The tendon pulls off the bone where it was reattached, at the repair site itself. The tendon substance remains intact and still reaches the bone.

Type 2 is failure at the medial row — the tendon tears through at the suture line where the medial anchors hold it, typically at or near the musculotendinous junction, while the tendon attached to the footprint remains healed to the bone. On MRI this looks paradoxical: a well-healed footprint with a defect medial to it.

The original description of these patterns compared techniques directly. Suture bridge better preserved the cuff tissue repaired to the insertion site than single-row did — but the retears that did occur with suture bridge were mainly at the musculotendinous junction, whereas single-row tended to fail directly at the footprint [16]. Imaging work after double-row repair found the same signature: complete tearing around the medial-row anchors with a well-repaired tendon on the footprint [17].

So the stronger construct did not abolish failure. It relocated it. Pooled across techniques, double-row and suture-bridge repairs increase the risk of medial cuff failure, and the authors note that technique modifications can reduce it [18].

Why Type 2 is the worse problem. Two reasons, and the second is the one that matters most.

Functionally, it does worse: in a series of 373 repairs with a retear rate of 15.6%, type 2 failures showed significantly inferior functional outcomes [19].

Anatomically, it is much harder to revise. A Type 1 failure leaves a tendon of full length that has simply detached — there is tissue to grasp, mobilise and reattach. A Type 2 failure has destroyed tendon substance at the medial row, so what remains is a shorter, often retracted tendon with the good, healed portion of it still fixed to the bone laterally. Revision then means working with less tendon than the original operation had, and pulling what is left further across a gap it cannot comfortably reach. In some cases direct repair is no longer possible at all, and the conversation moves to reconstruction, tendon transfer or reverse replacement.

That asymmetry is the reason technique modifications aimed specifically at reducing medial-row strain — avoiding over-tensioning, medial knot placement, and how the medial row is loaded — receive the attention they do. Not because Type 2 is common, but because it forecloses options.

One construct, named

To make this concrete, one specific construct: Dr Hirpara uses a medial row of all-suture anchors — the JuggerKnot (Zimmer Biomet) — placed at the articular margin, with the sutures carried laterally and secured by a lateral-row anchor, the Quattro (Zimmer Biomet), in the tuberosity. Other surgeons use different implants and different configurations, and this is offered as an illustration of the reasoning rather than as a recommendation.

The logic follows everything above. The medial row goes where bone quality is poorest and where sparing bone matters most, so a soft anchor with a small hole is used. The lateral row sits in better bone and takes the tensioning load, so a solid anchor is used. It is a construct assembled from the specific strengths of two generations of technology rather than a preference for either.


References for the advanced reading
  1. Imbergamo C, Sequeira S, Bano J, Rate WR, Gould H. Failure rates of suture anchor fixation versus transosseous tunnel techniques: a systematic review of biomechanical studies. Orthop J Sports Med. 2022;10(8).
  2. Na Y, Ren X, Wang Z, Shi Y, Wang J, Zhang L, et al. Arthroscopic transosseous anchorless versus suture anchor repair for rotator cuff tears: a systematic review and meta-analysis. BMC Musculoskelet Disord. 2026;27(1).
  3. Ozbaydar M, Elhassan B, Warner JJ. The use of anchors in shoulder surgery: a shift from metallic to bioabsorbable anchors. Arthroscopy. 2007;23(10):1124-6.
  4. Dhawan A, Ghodadra N, Karas V, Salata MJ, Cole BJ. Complications of bioabsorbable suture anchors in the shoulder. Am J Sports Med. 2011;40(6):1424-30.
  5. Barber FA, Herbert MA, Beavis RC, Barrera Oro F. Suture anchor materials, eyelets, and designs: update 2008. Arthroscopy. 2008;24(8):859-67.
  6. Ro K, Pancholi S, Son HS, Rhee YG. Perianchor cyst formation after arthroscopic rotator cuff repair using all-suture-type, bioabsorbable-type, and PEEK-type anchors. Arthroscopy. 2019;35(8):2284-92.
  7. Kim SH, Yang SH, Rhee S, Lee KJ, Kim HS, Oh JH. The formation of perianchor fluid associated with various suture anchors used in rotator cuff repair. Bone Joint J. 2019;101-B(12):1506-11.
  8. Mei Y, Jing L, He P, Wu W, Wang Z, Zhao F, et al. Period of formation of and factors associated with perianchor cysts around Healicoil and Footprint polyetheretherketone anchors after rotator cuff repair. J Shoulder Elbow Surg. 2026;35(1):e80-e86.
  9. Jeong HJ, Lee JS, Kim YK, Rhee S, Oh JH. Arthroscopic transosseous anchorless rotator cuff repair reduces bone defects related to peri-implant cyst formation. Clin Shoulder Elb. 2023;26(3):276-86.
  10. Ergün S, Akgün U, Barber FA, Karahan M. The clinical and biomechanical performance of all-suture anchors: a systematic review. Arthrosc Sports Med Rehabil. 2020;2(3):e263-e275.
  11. Feldman JJ, Ostrander B, Ithurburn MP, Fleisig GS, Tatum R, Ochsner MG, et al. The relationship between all-suture and solid medial-row anchors and patient outcomes in double-row rotator cuff repair. Orthop J Sports Med. 2024;12(8).
  12. Kim MS, Rhee SM, Cho NS. Perianchor cyst formation in all-suture anchor after rotator cuff repair: an evaluation of anchor insertion angle. J Shoulder Elbow Surg. 2022;31(9):1831-9.
  13. Boksh K, Haque A, Sharma A, Divall P, Singh H. Use of suture tapes versus conventional sutures for arthroscopic rotator cuff repairs: a systematic review and meta-analysis. Am J Sports Med. 2021;50(1):264-72.
  14. Hein J, Reilly JM, Chae J, Maerz T, Anderson K. Retear rates after arthroscopic single-row, double-row, and suture bridge rotator cuff repair at a minimum of 1 year of imaging follow-up: a systematic review. Arthroscopy. 2015;31(11):2274-81.
  15. Figueiredo JA, Sarmento M, Moura N, Gomes DS, Cartucho A. Knotted or knotless double-row rotator cuff repair retear rates: a systematic review and meta-analysis. JSES Rev Rep Tech. 2024;4(1):15-19.
  16. Cho NS, Yi JW, Lee BG, Rhee YG. Retear patterns after arthroscopic rotator cuff repair: single-row versus suture bridge technique. Am J Sports Med. 2009;38(4):664-71.
  17. Hayashida K, Tanaka M, Koizumi K, Kakiuchi M. Characteristic retear patterns assessed by magnetic resonance imaging after arthroscopic double-row rotator cuff repair. Arthroscopy. 2012;28(4):458-64.
  18. Bedeir YH, Schumaier AP, Abu-Sheasha G, Grawe BM. Type 2 retear after arthroscopic single-row, double-row and suture bridge rotator cuff repair: a systematic review. Eur J Orthop Surg Traumatol. 2018;29(2):373-82.
  19. Takeuchi Y, Sugaya H, Takahashi N, Matsuki K, Tokai M, Morioka T, et al. Repair integrity and retear pattern after arthroscopic medial knot-tying after suture-bridge lateral row rotator cuff repair. Am J Sports Med. 2020;48(10):2510-7.

Creative Commons BY-NC 4.0

CC Creative Commons licence
BY Attribution — you must credit the source
NC NonCommercial — not for commercial use

Attribution-NonCommercial 4.0 International


Creative Commons Corporation ("Creative Commons") is not a law firm and does not provide legal services or legal advice. Distribution of Creative Commons public licenses does not create a lawyer-client or other relationship. Creative Commons makes its licenses and related information available on an "as-is" basis. Creative Commons gives no warranties regarding its licenses, any material licensed under their terms and conditions, or any related information. Creative Commons disclaims all liability for damages resulting from their use to the fullest extent possible.

Using Creative Commons Public Licenses

Creative Commons public licenses provide a standard set of terms and conditions that creators and other rights holders may use to share original works of authorship and other material subject to copyright and certain other rights specified in the public license below. The following considerations are for informational purposes only, are not exhaustive, and do not form part of our licenses.

Considerations for licensors: Our public licenses are intended for use by those authorized to give the public permission to use material in ways otherwise restricted by copyright and certain other rights. Our licenses are irrevocable. Licensors should read and understand the terms and conditions of the license they choose before applying it. Licensors should also secure all rights necessary before applying our licenses so that the public can reuse the material as expected. Licensors should clearly mark any material not subject to the license. This includes other CC- licensed material, or material used under an exception or limitation to copyright. More considerations for licensors: wiki.creativecommons.org/Considerations_for_licensors

Considerations for the public: By using one of our public licenses, a licensor grants the public permission to use the licensed material under specified terms and conditions. If the licensor's permission is not necessary for any reason--for example, because of any applicable exception or limitation to copyright--then that use is not regulated by the license. Our licenses grant only permissions under copyright and certain other rights that a licensor has authority to grant. Use of the licensed material may still be restricted for other reasons, including because others have copyright or other rights in the material. A licensor may make special requests, such as asking that all changes be marked or described. Although not required by our licenses, you are encouraged to respect those requests where reasonable. More considerations for the public: wiki.creativecommons.org/Considerations_for_licensees


Creative Commons Attribution-NonCommercial 4.0 International Public License

By exercising the Licensed Rights (defined below), You accept and agree to be bound by the terms and conditions of this Creative Commons Attribution-NonCommercial 4.0 International Public License ("Public License"). To the extent this Public License may be interpreted as a contract, You are granted the Licensed Rights in consideration of Your acceptance of these terms and conditions, and the Licensor grants You such rights in consideration of benefits the Licensor receives from making the Licensed Material available under these terms and conditions.

Section 1 -- Definitions.

a. Adapted Material means material subject to Copyright and Similar Rights that is derived from or based upon the Licensed Material and in which the Licensed Material is translated, altered, arranged, transformed, or otherwise modified in a manner requiring permission under the Copyright and Similar Rights held by the Licensor. For purposes of this Public License, where the Licensed Material is a musical work, performance, or sound recording, Adapted Material is always produced where the Licensed Material is synched in timed relation with a moving image.

b. Adapter's License means the license You apply to Your Copyright and Similar Rights in Your contributions to Adapted Material in accordance with the terms and conditions of this Public License.

c. Copyright and Similar Rights means copyright and/or similar rights closely related to copyright including, without limitation, performance, broadcast, sound recording, and Sui Generis Database Rights, without regard to how the rights are labeled or categorized. For purposes of this Public License, the rights specified in Section 2(b)(1)-(2) are not Copyright and Similar Rights.

d. Effective Technological Measures means those measures that, in the absence of proper authority, may not be circumvented under laws fulfilling obligations under Article 11 of the WIPO Copyright Treaty adopted on December 20, 1996, and/or similar international agreements.

e. Exceptions and Limitations means fair use, fair dealing, and/or any other exception or limitation to Copyright and Similar Rights that applies to Your use of the Licensed Material.

f. Licensed Material means the artistic or literary work, database, or other material to which the Licensor applied this Public License.

g. Licensed Rights means the rights granted to You subject to the terms and conditions of this Public License, which are limited to all Copyright and Similar Rights that apply to Your use of the Licensed Material and that the Licensor has authority to license.

h. Licensor means the individual(s) or entity(ies) granting rights under this Public License.

i. NonCommercial means not primarily intended for or directed towards commercial advantage or monetary compensation. For purposes of this Public License, the exchange of the Licensed Material for other material subject to Copyright and Similar Rights by digital file-sharing or similar means is NonCommercial provided there is no payment of monetary compensation in connection with the exchange.

j. Share means to provide material to the public by any means or process that requires permission under the Licensed Rights, such as reproduction, public display, public performance, distribution, dissemination, communication, or importation, and to make material available to the public including in ways that members of the public may access the material from a place and at a time individually chosen by them.

k. Sui Generis Database Rights means rights other than copyright resulting from Directive 96/9/EC of the European Parliament and of the Council of 11 March 1996 on the legal protection of databases, as amended and/or succeeded, as well as other essentially equivalent rights anywhere in the world.

l. You means the individual or entity exercising the Licensed Rights under this Public License. Your has a corresponding meaning.

Section 2 -- Scope.

a. License grant.

1. Subject to the terms and conditions of this Public License, the Licensor hereby grants You a worldwide, royalty-free, non-sublicensable, non-exclusive, irrevocable license to exercise the Licensed Rights in the Licensed Material to:

a. reproduce and Share the Licensed Material, in whole or in part, for NonCommercial purposes only; and

b. produce, reproduce, and Share Adapted Material for NonCommercial purposes only.

2. Exceptions and Limitations. For the avoidance of doubt, where Exceptions and Limitations apply to Your use, this Public License does not apply, and You do not need to comply with its terms and conditions.

3. Term. The term of this Public License is specified in Section 6(a).

4. Media and formats; technical modifications allowed. The Licensor authorizes You to exercise the Licensed Rights in all media and formats whether now known or hereafter created, and to make technical modifications necessary to do so. The Licensor waives and/or agrees not to assert any right or authority to forbid You from making technical modifications necessary to exercise the Licensed Rights, including technical modifications necessary to circumvent Effective Technological Measures. For purposes of this Public License, simply making modifications authorized by this Section 2(a) (4) never produces Adapted Material.

5. Downstream recipients.

a. Offer from the Licensor -- Licensed Material. Every recipient of the Licensed Material automatically receives an offer from the Licensor to exercise the Licensed Rights under the terms and conditions of this Public License.

b. No downstream restrictions. You may not offer or impose any additional or different terms or conditions on, or apply any Effective Technological Measures to, the Licensed Material if doing so restricts exercise of the Licensed Rights by any recipient of the Licensed Material.

6. No endorsement. Nothing in this Public License constitutes or may be construed as permission to assert or imply that You are, or that Your use of the Licensed Material is, connected with, or sponsored, endorsed, or granted official status by, the Licensor or others designated to receive attribution as provided in Section 3(a)(1)(A)(i).

b. Other rights.

1. Moral rights, such as the right of integrity, are not licensed under this Public License, nor are publicity, privacy, and/or other similar personality rights; however, to the extent possible, the Licensor waives and/or agrees not to assert any such rights held by the Licensor to the limited extent necessary to allow You to exercise the Licensed Rights, but not otherwise.

2. Patent and trademark rights are not licensed under this Public License.

3. To the extent possible, the Licensor waives any right to collect royalties from You for the exercise of the Licensed Rights, whether directly or through a collecting society under any voluntary or waivable statutory or compulsory licensing scheme. In all other cases the Licensor expressly reserves any right to collect such royalties, including when the Licensed Material is used other than for NonCommercial purposes.

Section 3 -- License Conditions.

Your exercise of the Licensed Rights is expressly made subject to the following conditions.

a. Attribution.

1. If You Share the Licensed Material (including in modified form), You must:

a. retain the following if it is supplied by the Licensor with the Licensed Material:

i. identification of the creator(s) of the Licensed Material and any others designated to receive attribution, in any reasonable manner requested by the Licensor (including by pseudonym if designated);

ii. a copyright notice;

iii. a notice that refers to this Public License;

iv. a notice that refers to the disclaimer of warranties;

v. a URI or hyperlink to the Licensed Material to the extent reasonably practicable;

b. indicate if You modified the Licensed Material and retain an indication of any previous modifications; and

c. indicate the Licensed Material is licensed under this Public License, and include the text of, or the URI or hyperlink to, this Public License.

2. You may satisfy the conditions in Section 3(a)(1) in any reasonable manner based on the medium, means, and context in which You Share the Licensed Material. For example, it may be reasonable to satisfy the conditions by providing a URI or hyperlink to a resource that includes the required information.

3. If requested by the Licensor, You must remove any of the information required by Section 3(a)(1)(A) to the extent reasonably practicable.

4. If You Share Adapted Material You produce, the Adapter's License You apply must not prevent recipients of the Adapted Material from complying with this Public License.

Section 4 -- Sui Generis Database Rights.

Where the Licensed Rights include Sui Generis Database Rights that apply to Your use of the Licensed Material:

a. for the avoidance of doubt, Section 2(a)(1) grants You the right to extract, reuse, reproduce, and Share all or a substantial portion of the contents of the database for NonCommercial purposes only;

b. if You include all or a substantial portion of the database contents in a database in which You have Sui Generis Database Rights, then the database in which You have Sui Generis Database Rights (but not its individual contents) is Adapted Material; and

c. You must comply with the conditions in Section 3(a) if You Share all or a substantial portion of the contents of the database.

For the avoidance of doubt, this Section 4 supplements and does not replace Your obligations under this Public License where the Licensed Rights include other Copyright and Similar Rights.

Section 5 -- Disclaimer of Warranties and Limitation of Liability.

a. UNLESS OTHERWISE SEPARATELY UNDERTAKEN BY THE LICENSOR, TO THE EXTENT POSSIBLE, THE LICENSOR OFFERS THE LICENSED MATERIAL AS-IS AND AS-AVAILABLE, AND MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY KIND CONCERNING THE LICENSED MATERIAL, WHETHER EXPRESS, IMPLIED, STATUTORY, OR OTHER. THIS INCLUDES, WITHOUT LIMITATION, WARRANTIES OF TITLE, MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, NON-INFRINGEMENT, ABSENCE OF LATENT OR OTHER DEFECTS, ACCURACY, OR THE PRESENCE OR ABSENCE OF ERRORS, WHETHER OR NOT KNOWN OR DISCOVERABLE. WHERE DISCLAIMERS OF WARRANTIES ARE NOT ALLOWED IN FULL OR IN PART, THIS DISCLAIMER MAY NOT APPLY TO YOU.

b. TO THE EXTENT POSSIBLE, IN NO EVENT WILL THE LICENSOR BE LIABLE TO YOU ON ANY LEGAL THEORY (INCLUDING, WITHOUT LIMITATION, NEGLIGENCE) OR OTHERWISE FOR ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, CONSEQUENTIAL, PUNITIVE, EXEMPLARY, OR OTHER LOSSES, COSTS, EXPENSES, OR DAMAGES ARISING OUT OF THIS PUBLIC LICENSE OR USE OF THE LICENSED MATERIAL, EVEN IF THE LICENSOR HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH LOSSES, COSTS, EXPENSES, OR DAMAGES. WHERE A LIMITATION OF LIABILITY IS NOT ALLOWED IN FULL OR IN PART, THIS LIMITATION MAY NOT APPLY TO YOU.

c. The disclaimer of warranties and limitation of liability provided above shall be interpreted in a manner that, to the extent possible, most closely approximates an absolute disclaimer and waiver of all liability.

Section 6 -- Term and Termination.

a. This Public License applies for the term of the Copyright and Similar Rights licensed here. However, if You fail to comply with this Public License, then Your rights under this Public License terminate automatically.

b. Where Your right to use the Licensed Material has terminated under Section 6(a), it reinstates:

1. automatically as of the date the violation is cured, provided it is cured within 30 days of Your discovery of the violation; or

2. upon express reinstatement by the Licensor.

For the avoidance of doubt, this Section 6(b) does not affect any right the Licensor may have to seek remedies for Your violations of this Public License.

c. For the avoidance of doubt, the Licensor may also offer the Licensed Material under separate terms or conditions or stop distributing the Licensed Material at any time; however, doing so will not terminate this Public License.

d. Sections 1, 5, 6, 7, and 8 survive termination of this Public License.

Section 7 -- Other Terms and Conditions.

a. The Licensor shall not be bound by any additional or different terms or conditions communicated by You unless expressly agreed.

b. Any arrangements, understandings, or agreements regarding the Licensed Material not stated herein are separate from and independent of the terms and conditions of this Public License.

Section 8 -- Interpretation.

a. For the avoidance of doubt, this Public License does not, and shall not be interpreted to, reduce, limit, restrict, or impose conditions on any use of the Licensed Material that could lawfully be made without permission under this Public License.

b. To the extent possible, if any provision of this Public License is deemed unenforceable, it shall be automatically reformed to the minimum extent necessary to make it enforceable. If the provision cannot be reformed, it shall be severed from this Public License without affecting the enforceability of the remaining terms and conditions.

c. No term or condition of this Public License will be waived and no failure to comply consented to unless expressly agreed to by the Licensor.

d. Nothing in this Public License constitutes or may be interpreted as a limitation upon, or waiver of, any privileges and immunities that apply to the Licensor or You, including from the legal processes of any jurisdiction or authority.


Creative Commons is not a party to its public licenses. Notwithstanding, Creative Commons may elect to apply one of its public licenses to material it publishes and in those instances will be considered the “Licensor.” The text of the Creative Commons public licenses is dedicated to the public domain under the CC0 Public Domain Dedication. Except for the limited purpose of indicating that material is shared under a Creative Commons public license or as otherwise permitted by the Creative Commons policies published at creativecommons.org/policies, Creative Commons does not authorize the use of the trademark "Creative Commons" or any other trademark or logo of Creative Commons without its prior written consent including, without limitation, in connection with any unauthorized modifications to any of its public licenses or any other arrangements, understandings, or agreements concerning use of licensed material. For the avoidance of doubt, this paragraph does not form part of the public licenses.

Creative Commons may be contacted at creativecommons.org.