Patients › Shoulder

Robotic-Assisted Shoulder Replacement

Updated Oct 202621 citations
Illustration: shoulder

What is it

Robotic-assisted shoulder replacement is an anatomic or reverse shoulder replacement in which a robotic arm helps carry out the plan made before surgery. The operation itself does not change: it treats a worn joint (arthritis) and, with a reverse replacement, a shoulder with a badly damaged rotator cuff.

Before the operation, a CT scan of your shoulder is turned into a three-dimensional plan that sets the size, angle and depth of each part of the new joint. During the operation, the robotic arm positions the cutting guide for the top of the arm bone and holds the reamer that shapes the socket, at the angles and depth in that plan, then holds still while your surgeon makes the cut or shapes the socket. The robot does not operate on its own or make decisions; your surgeon performs every step.

What it's made of

The system has four parts: a robotic arm on a wheeled base beside the operating table; a camera that tracks small reflective markers; the tracking pins those markers sit on, placed in the arm bone and in the coracoid (a bony knob at the front of the shoulder blade) through the same incision as the operation; and the plan made from your CT scan. The pins come out at the end of the operation, and nothing robotic stays in your body. The implants are those of an anatomic or reverse shoulder replacement, described on those pages.

How your surgeon uses it

Dr Kieran Hirpara, an upper-limb surgeon at Mater Private Hospital Rockhampton, uses the ROSA robot (Zimmer Biomet) with the Zimmer Biomet Comprehensive shoulder replacement, planned with the Signature ONE planning software. We already plan every shoulder replacement on a CT scan, so a robotic operation needs no extra scan. We use robotic assistance routinely for both anatomic and reverse replacements. If the robot has a problem during the operation, we continue with patient-specific guides made from the same plan, and with standard instruments if needed. Other surgeons use other systems, such as Mako, or other aids such as navigation or patient-specific guides, and many use none.

What to expect

Your anaesthetic, hospital stay, sling and rehabilitation are the same as for any shoulder replacement, and the anatomic and reverse replacement pages describe them in detail. The robot adds no extra wounds or appointments. The operation may take a little longer at first while the theatre team becomes familiar with the system.

What the evidence says

The clearest finding so far is about accuracy. With robotic assistance, the position of the implant matches the plan more consistently, and the occasional large miss that can happen when parts are placed by hand is removed [1]. That is what the robot has been shown to do.

It has not yet been shown to change how the shoulder feels or how long the implant lasts. No study has shown better pain relief, better function or a longer-lasting implant in the shoulder with robotic help [2]. Related technologies point the same way. Navigation, which guides the surgeon's hand rather than holding the instrument, placed the socket more accurately [3], with results at two years at least as good as standard instruments and no extra complications [4]. Patient-specific guides showed no major differences in results over the long term [5].

The evidence is still building. Most published accuracy studies used a different robot, and outcome studies of the ROSA shoulder system are still to come [6]. Longer follow-up is needed before anyone can say whether greater accuracy changes how long a shoulder replacement lasts [7]. In robotic hip and knee replacement, where most of the experience lies, reported device problems have been uncommon, and the rate has fallen over time [8].

In short, the robot helps carry out the plan precisely and consistently. Whether that leads to a shoulder that works better or lasts longer is a question research has not yet answered.

In more depth

Advanced reading: the deeper science (optional)

This section goes further than you need for your own treatment decisions. Robotic shoulder replacement is worth the extra reading because what it improves and what people expect it to improve are two different things. It reliably helps the surgeon carry out the plan; whether that changes how the shoulder feels and lasts is not yet known.

Collaborative, not autonomous

Shoulder replacement robots are collaborative: the arm moves an instrument into the planned position and then locks still, and the surgeon makes the cut or shapes the socket [1]. The steps ROSA guides are described above; the rest of the operation is done by hand in the usual way [2]. The robot does not decide anything.

What it improves: the misses, not the average

The clearest evidence is about accuracy against the plan. Almost all of it so far comes from the Mako robot (Stryker), which shapes the socket with a burr; ROSA holds a conventional reamer instead, and its own accuracy studies are still to come. In a laboratory study on 3D-printed shoulder blades, socket preparation with the Mako arm matched the plan more closely than both hand preparation and patient-specific guides: the angle of the socket was off by about 1° on average with the robot, against 9° by hand and 4° with guides [3]. In patients, one surgeon's Mako series of 103 reverse replacements found the average accuracy was similar with and without the robot, but the spread was narrower: no robotic case was more than 10° away from the plan, against 4 of 42 (10%) done conventionally [4].

That is the real argument for the technology. Most hand-placed implants are already close to the plan; the robot's contribution is to remove the occasional large miss. Whether removing those outliers changes how long an implant lasts is the question no study has yet answered.

What has not been shown yet

No study has yet shown that robotic shoulder replacement gives better pain relief, better movement or a longer-lasting implant than a well-performed conventional operation. The technologies that came before it tell a consistent story. Navigation made the socket angle about 6° more accurate on average across pooled studies, but the authors questioned whether that difference matters clinically [5]. A meta-analysis of 537 patients found no difference in patient-reported results or complications with patient-specific guides [6], and a long-term comparison of reverse replacements found no major clinical differences with them either [7]. Across hip, knee and shoulder, reviews of trials and national registries have found robotic surgery reliably improves positioning but has not consistently improved outcomes, complication rates or implant survival [8].

That is not the same as "it makes no difference". It means the benefit, if there is one, is in reducing uncommon problems, and that takes large numbers and long follow-up to prove.

The learning curve, theatre time and the machine itself

  • Learning curve. In an Australian Mako series, the workload measured for the surgical team levelled off after 3 to 5 robotic cases [9].
  • Theatre time. In a Mako series of 77 reverse replacements, robotic cases took about the same time on average as conventional ones but were more predictable [10].
  • Technology failure. In a US national database of robotic hip and knee replacement, device problems were reported in 0.3 to 0.4% of operations, with delays in 0.13 to 0.22%, and the rate fell year on year [11].

Expectations

In a survey of 615 American adults, 67% believed robotic shoulder replacement gives better results and 54% a faster recovery, neither of which has been shown. Asked to choose, 62% would rather have a more experienced surgeon who does not use a robot than a less experienced one who does [12]. Experts writing on consent for robotic shoulder surgery make the same point: the robot is a tool that helps carry out a plan, it does not replace the surgeon's judgement, and its long-term benefits are not yet known [13].

What this means for you

  • The robot changes how precisely the plan is carried out, not what the operation is.
  • Its proven effect is consistency: fewer large departures from the plan.
  • Better pain, function or implant life have not been shown, though they may follow.
  • It is one of several ways to achieve an accurate plan; navigation and patient-specific guides do similar jobs, and many surgeons use none of them.

References for the advanced reading
  1. Robotic assistance improves reproducibility of planned glenoid baseplate inclination in reverse shoulder arthroplasty. *International Orthopaedics*. 2026. 10.1007/s00264-026-06969-y
  2. Technological innovations in shoulder replacement: current concepts and the future of robotics in total shoulder arthroplasty. *Journal of Shoulder and Elbow Surgery*. 2023. 10.1016/j.jse.2023.04.022
  3. Accuracy of placement of the glenoid component in reversed shoulder arthroplasty with and without navigation. *Journal of Shoulder and Elbow Surgery*. 2011. 10.1016/j.jse.2010.07.014
  4. Two-year clinical outcomes and complication rates in anatomic and reverse shoulder arthroplasty implanted with Exactech GPS intraoperative navigation. *Journal of Shoulder and Elbow Surgery*. 2023. 10.1016/j.jse.2023.05.021
  5. Long-term outcomes after reverse shoulder arthroplasty show no improvements with patient-specific instrumentation. *JSES International*. 2026. 10.1016/j.jseint.2025.101456
  6. Robotic assisted reverse total shoulder arthroplasty: narrative review and surgical technique of humeral and glenoid preparation. *JSES Reviews, Reports, and Techniques*. 2026. 10.1016/j.xrrt.2026.100708
  7. Computer-Assisted Navigation in Shoulder Arthroplasty: A Narrative Review. *Journal of Clinical Medicine*. 2025. 10.3390/jcm14082763
  8. How Often Does Technology Fail in Robotic-Assisted Arthroplasty? A Comprehensive Analysis of a United States Food and Drug Administration Database. *The Journal of Arthroplasty*. 2025. 10.1016/j.arth.2024.09.005
  9. Sanchez-Sotelo J. Robot-assisted shoulder arthroplasty. JSES Int. 2025;9(3):974-80.
  10. Marigi EM, Lopez CE, Marigi IM, Duquin TR, Sperling JW. Robotic assisted reverse total shoulder arthroplasty: narrative review and surgical technique of humeral and glenoid preparation. JSES Rev Rep Tech. 2026;6(2):100708.
  11. Athwal GS, Nelson A, Antuna S, Ponce B, Mighell M, St Pierre P, et al. Glenoid preparation in reverse shoulder arthroplasty: robotic arm-assisted preparation compared to manual preparation and patient-specific guides. J Shoulder Elbow Surg. 2025;34(8):2022-30.
  12. Singh D, Hamawandi M, Menendez ME. Robotic assistance improves reproducibility of planned glenoid baseplate inclination in reverse shoulder arthroplasty. Int Orthop. 2026.
  13. Sadoghi P, Vavken J, Leithner A, Vavken P. Benefit of intraoperative navigation on glenoid component positioning during total shoulder arthroplasty. Arch Orthop Trauma Surg. 2014;135(1):41-7.
  14. Daher M, Parmar T, Boufadel P, Fares MY, Khalil W, Horneff JG, et al. Patient-specific instrumentation in primary total shoulder arthroplasty: a meta-analysis of clinical outcomes. Clin Shoulder Elb. 2025;28(2):129-36.
  15. Dordain F, Nourissat G, Ohl X, Housset V. Long-term outcomes after reverse shoulder arthroplasty show no improvements with patient-specific instrumentation. JSES Int. 2026;10(2):101456.
  16. Fraval A, Rajgopal A, Bozic KJ, Bhandari M, Parvizi J. Robotic Arthroplasty in Context. J Bone Joint Surg Am. 2026.
  17. Boekel P, Grant A, Doma K, Morse L. The learning curve associated with robotic-assisted shoulder arthroplasty: A feasibility study. Shoulder Elbow. 2026.
  18. Frederickson M, Singh D, Menendez ME. Robotic assistance reduces operative time variability in reverse shoulder arthroplasty: a retrospective single-surgeon study. Int Orthop. 2026.
  19. Costello JP, Constantinescu DS, Chen JA, Cai NA, Vigdorchik JM, Silva Iacobelli DA. How Often Does Technology Fail in Robotic-Assisted Arthroplasty? A Comprehensive Analysis of a United States Food and Drug Administration Database. J Arthroplasty. 2025;40(3):584-91.
  20. Zabiba A, Menendez ME. Public Perceptions of Robotic-Assisted Shoulder Arthroplasty in the United States. J Shoulder Elbow Surg. 2026.
  21. Menendez ME, Sanchez-Sotelo J. Informed Consent for Robotic-Assisted Reverse Shoulder Arthroplasty. J Bone Joint Surg Am. 2026.
Evidence & references

This is the clinical evidence summary written for health professionals. It is technical, and it lists the research this page was built from. You do not need to read it to understand your treatment or to make a decision about it.

Anatomy & Pathophysiology

Bony Anatomy

  • The proximal humerus comprises four main parts: the humeral head, greater tuberosity (GT), lesser tuberosity (LT), and humeral shaft [3].
  • The articular head of the humerus is spherical with a diameter of 37 to 57 mm [3].
  • The most superior portion of the articular surface of the humeral head averages 8 mm above the greater tuberosity [3].
  • Humeral version averages 29.8 degrees, with a range of 10 to 55 degrees [3].
  • The humeral head is inclined approximately 130 degrees with respect to the humeral shaft [3].
  • The neck-shaft angle measures an average of 135 degrees [4].
  • The humeral head is retroverted an average of 30 degrees [4].
  • The humeral head averages 19° of retroversion and 41° of inclination (neck-shaft angle) [6].
  • The articular surface of the humeral head is essentially spherical, with an arc of approximately 160 degrees covered by articular cartilage [15].
  • The radius of curvature of the humeral head is approximately 25 mm and is slightly larger in men than in women [15].
  • The average neck-shaft angle is 45 degrees (±5 degrees), with a range of 30 to 50 degrees [15].
  • Arthritic shoulders have a flatter neck-shaft angle close to 50 degrees [15].
  • The superior margin of the humeral head articular surface is normally superior to the top of the greater tuberosity by 8 to 10 mm [15].
  • The distance from the lateral base of the coracoid process to the lateral margin of the greater tuberosity is called the lateral humeral offset [15].
  • A significant decrease in lateral humeral offset reduces the lever arms for the deltoid and supraspinatus muscles, weakening abduction and impairing function [15].
  • A significant increase in lateral humeral offset causes excessive tension on the soft tissues ("overstuffing" of the joint), resulting in loss of motion and likely accelerating polyethylene wear [15].
  • Humeral articular malposition of more than 4 mm leads to increased subacromial contact [15].
  • An offset of 8 mm in any direction significantly decreases passive range of motion [15].
  • Proximal humeral retroversion is highly variable, ranging from 0 to 55 degrees depending on the method used for measurement [15].
  • The glenoid is a convex structure of shallow depth shaped like an inverted pear [3].
  • The glenoid cavity is a shallow socket, approximately one third the size of the humeral head [4].
  • The subchondral bone of the glenoid is relatively flat, and the articular concavity is augmented by cartilage and a circumferential labrum [6].
  • The glenoid averages 5° of retroversion in relation to the axis of the scapular body [6].
  • The normal position of the glenoid surface in relation to the axis of the scapular body ranges from 2 degrees of anteversion to 7 degrees of retroversion [15].
  • The glenoid articular surface radius of curvature is 2 to 3 mm larger than that of the humeral head [15].
  • The glenoid diameter ranges from 18-30 mm superior anteroposterior, 21-35 mm inferior anteroposterior, and 30-48 mm superoinferior [15].
  • The glenoid inclination averages 4.2 degrees, with a range of –7 to 20 degrees [15].
  • The glenoid version averages 1.5 degrees retroversion, with a range of 10.5-9.5 degrees anteversion [15].
  • The glenoid surface area is 4-6 mm [15].
  • The humeral head surface area is 11-19 mm [15].
  • The glenoid cartilage thickness is 2.16 mm [15].
  • The humeral head cartilage thickness is 1.44 mm [15].
  • The glenoid radius of curvature is 22-28 mm [15].
  • The humeral head radius of curvature is 23-28 mm, smaller in women than men [15].
  • The medial (coronal) humeral offset is 4-14 mm [15].
  • The posterior (transverse) humeral offset is –2 to 10 mm [15].
  • The head-shaft angle is 30-55 degrees [15].
  • The scapula is attached to the axial skeleton by the clavicle, specifically via the acromioclavicular (AC) and sternoclavicular (SC) joints [5].
  • The scapula is separated from the chest wall by thin gliding fibro-fatty tissue, allowing smooth excursion over the chest wall [5].
  • The basic part of the scapula is the body, which is triangular when viewed anteroposteriorly with its base situated superiorly and its apex inferiorly [5].
  • The glenoid is connected with the flat body of the scapula by the scapular neck [5].
  • The hook-shaped coracoid process curves forwards from the superior surface of the scapular neck [5].
  • The scapular spine ends in a flattened bony process, the acromion, which curves forwards [5].
  • The highest concentration of bony mass in the scapula is found in the glenoid, the scapular neck (including the base of the coracoid process), and the lateral border of the scapular body [5].
  • Two bony pillars transmit compressive forces from the glenoid fossa: the lateral pillar and the spinal pillar [5].
  • The lateral pillar connects the inferior border of the glenoid with the inferior angle [5].
  • The spinal pillar arises from the central part of the glenoid and continues medially to become part of the base of the scapular spine [5].
  • The triangle formed by the two pillars and the medial border constitutes the biomechanical body of the scapula [5].
  • The weakest bone in the scapula is located primarily in the central part of the biomechanical body, i.e., in the infraspinous fossa [5].
  • The weakest area of the circumference of the biomechanical body of the scapula is the spinomedial angle, where the scapular spine connects to the medial border [5].
  • The clavicle is the first bone to ossify at the fifth week of gestation and is the only long bone to ossify by intramembranous ossification [6].
  • The medial (sternal) epiphysis of the clavicle is the last ossification center to fuse, at age 20 to 25 years [6].
  • The primary blood supply to the clavicle is periosteal, and no nutrient artery is present [6].
  • The scapula has only one true diarthrodial articulation, the acromioclavicular (AC) joint [6].
  • Normal shoulder motion is approximately two-thirds glenohumeral and one third scapulothoracic [6].
  • Ossification of the scapular body begins at the eighth week of gestation [6].
  • The acromion has three ossification centers: the metacromion (base), the mesoacromion (middle), and the preacromion (tip) [6].
  • Failure of fusion of the acromial ossification centers results in os acromiale [6].
  • The coracobrachialis muscle and the short head of the biceps tendon originate from the coracoid process [6].
  • The pectoralis minor muscle inserts onto the medial coracoid process [6].
  • The proximal humerus has three centers of ossification: the humeral head (4 to 6 months), the greater tuberosity (1 to 3 years), and the lesser tuberosity (3 to 5 years) [6].
  • The proximal humeral ossification centers fuse to the shaft at age 17 to 20 years [6].
  • The humeral head is retroverted 30 degrees relative to the transepicondylar axis of the humerus [17].
  • Head height is approximately 5.6 cm above the superior border of the pectoralis major tendon [17].
  • The anatomic neck is located directly below the humeral head and serves as an attachment for the shoulder capsule [17].
  • The surgical neck is more distal than the anatomic neck and is more often involved in fractures [17].
  • The scapula spans second through seventh ribs and serves as an attachment for 17 muscles [17].
  • The scapula is anteverted on the chest wall approximately 30 degrees relative to the body [17].
  • The glenoid is retroverted approximately 5 degrees relative to the scapular body [17].
  • Os acromiale is incomplete fusion of secondary ossification centers, most commonly between the mesoacromion and meta-acromion [17].
  • The coracoacromial ligament contributes to anterosuperior stability in rotator cuff deficiency and should be preserved with irreparable cuff tears to prevent anterosuperior escape [17].
  • The acromial branch of the thoracoacromial artery runs on the medial aspect of the coracoacromial ligament [17].
  • The coracoacromial ligament is the arthroscopic landmark for a complete release of the rotator interval for adhesive capsulitis [17].
  • The clavicle is the first bone in the body to ossify at 5 weeks gestation and last to fuse with the medial epiphysis at 25 years of age [17].
  • Fracture of the clavicle is the most common musculoskeletal birth injury [17].

Vascular Supply

  • The proximal humerus receives its blood supply from the anterior and posterior humeral circumflex branches from the third division of the axillary artery [3].
  • The posterior humeral circumflex artery travels with the axillary nerve, enters the quadrilateral space posteriorly, and anastomoses with a branch of the anterior circumflex to supply the posterior cuff [3].
  • The anterior humeral circumflex artery (AHCA) arises from the axillary artery at the inferior border of the subscapularis [3].
  • The AHCA provides vascular inflow to the humeral head by way of its terminal anterolateral branch known as the artery of Laing (also known as the arcuate artery) [3].
  • The ascending branch of the AHCA courses parallel to the lateral aspect of the long head biceps tendon and enters the humeral head at the interface of the bicipital groove and greater tuberosity [3].
  • Injury to the arcuate artery may result in osteonecrosis of the humeral head [3].
  • Additional extraosseous collateral branches can permit humeral head perfusion despite complete ligation of the arcuate artery [3].
  • The major blood supply to the humeral head is through the ascending branch of the anterior humeral circumflex artery, which penetrates the head at the bicipital groove and becomes the arcuate artery [4].
  • The anterolateral ascending branch of the anterior humeral circumflex artery provides the primary blood supply to the humeral head [6].
  • The terminal intraosseous portion of the anterior humeral circumflex artery enters at the proximal aspect of the intertubercular groove as the arcuate artery [6].
  • The anterior and posterior humeral circumflex arteries provide a rich blood supply to the proximal humerus [10].
  • Quantitative assessment has shown that 64% of the humeral head blood supply arises from the posterior humeral circumflex artery [10].
  • The primary blood supply to the clavicle is periosteal, and there is no nutrient blood supply [12].

Muscles and Tendons

  • The greater tuberosity serves as the attachment site for the supraspinatus, infraspinatus, and teres minor tendons of the rotator cuff [3].
  • The lesser tuberosity serves as the attachment site for the subscapularis tendon [3].
  • The bicipital groove lies between the greater and lesser tuberosities and serves as a pathway for the long head of the biceps [3].
  • The distal aspect of the bicipital groove is internally rotated with respect to the proximal portion [3].
  • The rotator cuff consists of four muscles: the subscapularis, supraspinatus, infraspinatus, and teres minor [4].
  • The teres major is not a rotator cuff muscle [4].
  • The cuff muscles serve as depressors of the humeral head to allow the deltoid to efficiently abduct the humerus [4].
  • The infraspinatus and teres minor are external rotators, while the subscapularis is an internal rotator of the humerus [4].
  • The deltoid and pectoralis major muscles, along with the rotator cuff, cause predictable displacement of fractures around the proximal humerus [4].
  • The rotator cuff is a sheet of conjoined tendons closely applied over the shoulder capsule and inserting mainly into the greater tuberosity of the humerus, with the subscapularis inserted into the lesser tuberosity [11].
  • The rotator cuff has an important function in stabilizing the head of the humerus by pulling it firmly into the glenoid whenever the deltoid lifts the arm forwards or sideways [11].
  • The rotator cuff must act simultaneously and synergistically with the deltoid muscle for normal function [15].
  • The extrinsic muscles primarily control movement of the scapula and include the rhomboids, levator scapulae, trapezius, and serratus anterior [16].
  • The intrinsic muscles control the glenohumeral joint and include the rotator cuff muscles, deltoid, pectoralis major, teres major, latissimus dorsi, and biceps brachii [16].
  • The tendinous insertions of the rotator cuff muscles, the articular capsule, the coracohumeral ligament, and the glenohumeral ligament complex blend into a confluent sheet before insertion into the humeral tuberosities [16].
  • The tendons of the infraspinatus and supraspinatus muscles join approximately 15 mm proximal to their insertion and cannot be readily separated by blunt dissection [16].
  • The infraspinatus and teres minor fuse near their musculotendinous junctions [16].
  • The supraspinatus and subscapularis tendons join as a sheath that surrounds the biceps tendon at the entrance of the bicipital groove [16].
  • The roof of the biceps sheath consists of a portion of the supraspinatus tendon, and a sheet of the subscapularis tendon forms the floor [16].
  • The coracohumeral ligament is a thick band of fibrous tissue extending from the coracoid process along the surface of the capsule to the tuberosities between the supraspinatus and subscapularis tendons [16].
  • The coracohumeral ligament is deep to the tendinous insertion of the cuff and blends with the capsule and supraspinatus tendon to form part of the roof of the biceps sheath [16].
  • The deltoid has anatomic subdivisions including acromial, clavicular, and scapular portions [8].
  • The deltoid is a well-defined muscle that is easily resected but extremely difficult to reconstruct functionally after proximal humeral resections for bone tumors [8].
  • The subscapularis originates from the anterior scapula and inserts anteriorly onto the lesser tuberosity [10].
  • The greater tuberosity provides attachment superiorly and posteriorly for the supraspinatus, infraspinatus, and teres minor, all of which originate from the posterior scapula [10].
  • The deltoid forward flexes and abducts the shoulder and courses from the clavicle and acromion superiorly, coalescing into a common tendinous insertion onto the lateral upper third of the humeral shaft [10].
  • The pectoralis major powers adduction and internal rotation due to its tendinous insertion anteriorly onto the lateral wall of the bicipital groove [10].
  • The pectoralis major forms the roof of the distal continuation of the bicipital tunnel, which is a closed space that extends proximally to the glenohumeral joint [10].
  • The infraspinatus and teres minor muscles are more effective in their roles as depressors and external rotators of the humeral head due to evolutionary changes in the vector of muscle pull from the axillary border of the scapula to the glenoid fossa [13].
  • The acromion has enlarged over time to reflect the increasing role of the deltoid muscle in shoulder function [13].
  • The broader attachment of the deltoid on the acromion and its more distal insertion on the humerus have increased its mechanical advantage in shoulder motion [13].

Ligaments and Stability

  • The glenohumeral joint depends on static and dynamic stabilizers for movement and stability, especially the rotator cuff [15].
  • The rotator cuff stabilizes the glenohumeral joint while allowing greater freedom of motion and fixes the fulcrum of the upper extremity against which the deltoid can contract and elevate the humerus [15].
  • The shoulder joint is composed of four articulations: the sternoclavicular, acromioclavicular, glenohumeral, and scapulothoracic [16].
  • The bony anatomy contributes little to stability and has been compared with

Investigations

Plain Radiography

  • The purpose of shoulder imaging is to establish diagnosis, determine pathoanatomy severity, assist in surgical planning, and illustrate the condition to the patient [2].
  • Standardized plain films are almost always sufficient to garner the information needed for care [2].
  • CT scans may offer increased precision in measuring glenoid version, but this precision does not necessarily improve the quality of surgery or clinical outcome [2].
  • The first key radiographic view is the anteroposterior (AP) view in the plane of the scapula, taken so the x-ray beam passes through the glenohumeral joint [2].
  • The AP view in the plane of the scapula shows the superoinferior position of the humeral head relative to the glenoid, presence of osteophytes, joint space narrowing, degree of medial humeral displacement, bone quality, loose bodies, and humeral head collapse or deformity [2].
  • The second key radiographic view is the axillary view taken with the arm in the functional position of elevation in the plane of the scapula [2].
  • The axillary view is oriented so that both the spinoglenoid notch and the scapular neck are visible [2].
  • The axillary view demonstrates glenohumeral relationships in the functional position of elevation, referred to as the "truth view" [2].
  • CT scans have the disadvantage of being taken with the arm in the adducted position, unlike the axillary truth view [2].
  • The standardized axillary view shows the thickness of the cartilage space, relative positions of the humeral head and glenoid, presence of osteophytes, degree of osteopenia, and extent of bony deformity and erosion [2].
  • Joint space narrowing is most evident on the axillary truth view because arthritis usually involves the central aspect of the humeral head [2].
  • The axillary truth view can show posterior subluxation or "functional decentering" that is not evident in images taken with the arm at the side [2].
  • The degree of posterior subluxation can be measured by the position of the center of the humeral head in relation to the plane of the scapula [2].
  • The degree of posterior subluxation can be measured by the position of the center of the humeral head in relation to the glenoid face [2].
  • The degree of posterior subluxation can be measured by the point of contact of the humeral articular surface on the glenoid articular surface [2].
  • The point of contact measurement reflects the degree of centering of the net humeral joint reaction force on the glenoid [2].
  • Malcentering of the joint reaction force leads to posterior instability, posterior glenoid wear, and "rocking horse" loosening of prosthetic glenoid components [2].
  • At least two X-ray views should be obtained: an anteroposterior in the plane of the glenoid and an axillary projection with the arm in abduction [20].
  • The standard shoulder series should include a true AP view in the scapular plane, an AP view, an axillary view, and a scapular Y view [28].
  • The true AP view in the scapular plane visualizes the anterior greater tuberosity in profile and can reveal proximal humeral migration [28].
  • The AP view with the arm in internal rotation visualizes the posterior aspect of the greater tuberosity and the lesser tuberosity in profile [28].
  • The axillary view is necessary for evaluating glenohumeral joint instability and determining humeral head position in the glenoid fossa [28].
  • The axillary view may detect occult, locked posterior shoulder dislocation in a patient who exhibits a lack of passive external rotation [28].
  • The scapular Y view provides visualization of the coracoacromial arch and can reveal coracoacromial spurs associated with rotator cuff pathology [28].
  • The scapular Y view is a reliable alternative for evaluating glenohumeral subluxation and dislocation [28].
  • The acromiohumeral distance is normally 7 to 14 mm [28].
  • The width of the glenohumeral joint space should be symmetric superiorly and inferiorly [28].
  • The coracoclavicular distance is normally 1.1 to 1.3 cm [28].
  • Neer classified acromial morphology as type I (flat), type II (curved), and type III (hooked) [28].
  • Type III acromial morphology has a correlation with the presence of rotator cuff disease, but no direct causal relationship has been demonstrated [28].
  • The Neer trauma series consists of an AP view, a lateral view in the scapular plane, and a Velpeau modified axillary view [22].
  • The lateral radiograph in the scapular plane is the tangential Y-view of the scapula [22].
  • The combination of the three views in the Neer trauma series allows evaluation of the shoulder joint in three separate perpendicular planes [22].
  • The axillary view is important for evaluating the glenoid articular surface and the relationship of the humeral head anteriorly and posteriorly [22].
  • Conventional radiographs are appropriate for patients presenting with shoulder pain with any history of trauma, dislocation, night pain, or chronic shoulder pain [28].
  • Arthritis, calcific tendinitis, and osteolysis of the distal clavicle can be observed on plain radiograph [27].

Computed Tomography (CT)

  • CT is helpful for planning fracture surgery and shoulder joint replacement [20].
  • CT imaging is frequently used to evaluate fractures of the shoulder, assess bony lesions in recurrent instability cases, or for preoperative templating for shoulder arthritis [27].
  • CT with three-dimensional reconstructions is the advanced imaging study of choice for determining the extent of glenoid bone loss in the setting of shoulder instability [28].
  • Three-dimensional sagittal CT scans allow for optimal evaluation of displaced glenoid fractures [28].
  • Three-dimensional reconstructions can reveal fine details of shoulder anatomy, but this additional information rarely changes the planning or conduct of the arthroplasty [2].

Magnetic Resonance Imaging (MRI)

  • MRI is useful to identify osteonecrosis of the humeral head, or a bone tumour [20].
  • MRI can identify labral tears and rotator cuff tears, although accuracy for these is enhanced by combining the scan with arthrography [20].
  • MRI is the modality of choice for evaluating the rotator cuff, biceps, and subacromial/subdeltoid bursa [27].
  • T1-weighted MRI can reveal Hill-Sachs lesions [27].
  • T2-weighted MRI provides better visualization of full thickness rotator cuff tears [27].
  • Traditional MRI is utilized for evaluation of soft tissues with high contrast and spatial resolution [25].
  • MR accuracy in identifying labral and rotator cuff tears in the literature ranges from 70% to 100% [25].
  • Acquired multi-planar imaging allows for the detailed evaluation of the glenoid, labrum, joint capsule, and rotator cuff in different planes [25].
  • With standard MRI or MRA, the shoulder is routinely positioned in neutral or partial external rotation [25].
  • Abduction and external rotation (ABER) of the arm is an alternative position utilized to increase the sensitivity and specificity for detecting anteroinferior labroligamentous injury [25].
  • Limited range of motion or pain may prohibit patients from performing the ABER provocative maneuver [25].
  • Full routine MRI or MRA examination had similar accuracy as the ABER sequence in evaluating the anteroinferior labral–ligamentous complex [25].
  • The sensitivity of MRA with the ABER position for detecting anteroinferior labral lesions was significantly higher than that of MRA in neutral position [25].
  • MRA with the ABER position is more effective in identifying Perthes lesions than MRA in neutral position [25].
  • MRAs can demonstrate a patulous capsule on coronal, sagittal, and axial imaging in patients with multidirectional instability [25].
  • MRAs can be helpful in evaluating lesions of the rotator interval and other associated findings that may affect the surgical plan [25].
  • Glenoid dysplasia, increased capsular cross-sectional area, and increased glenoid retroversion have been found to be associated with increased posterior labral tears and symptomatic instability [25].
  • Glenoid retroversion was significantly increased in patients with symptomatic posterior labral tears [25].
  • There was no significant association between instability and increased humeral head subluxation in patients with multidirectional instability [25].
  • The diagnosis of multidirectional instability is a clinical one, and the need for expensive or invasive imaging should be weighed against the information gained [25].

Arthrography

  • Arthrography involves injection of contrast agent in conjunction with either an MRI or CT scan to enhance imaging of the joint [27].
  • MR arthrography (MRA) refers to MRI of a joint that has been injected with an intra-articular contrast agent such as diluted gadolinium or saline solution [25].
  • The contrast material is injected prior to MRI by fluoroscopic or ultrasound guidance under strict aseptic technique [25].
  • By distending the joint capsule, the cartilage, ligaments, and labrum are outlined with contrast, increasing the sensitivity for detecting tears and other lesions [25].
  • In the acute dislocation setting, a joint effusion with distension of the joint may outline structures similarly, making the arthrogram unnecessary [25].
  • MRA has proven utility by increasing both sensitivity and specificity in detecting injuries to the capsulolabral–ligamentous complex compared to traditional MRI [25].
  • A meta-analysis of 6 studies including 4,667 shoulders found greater diagnostic test accuracy for MRA over MRI in the detection of glenoid labral lesions [25].
  • MRA sensitivity for detecting glenoid labral lesions was 88% and specificity was 93% [25].
  • MRI sensitivity for detecting glenoid labral lesions was 76% and specificity was 87% [25].
  • MR arthrography is considered the benchmark for evaluation of labral tears [27].
  • MR arthrography is rarely indicated for evaluation of rotator cuff pathology [27].
  • When MRI or MR arthrography is contraindicated, such as in patients with pacemakers or vascular clips, CT arthrography is indicated [27].

Ultrasonography

  • Ultrasonography is a simple and accurate test for identifying rotator cuff tears and calcific tendinitis [20].
  • Ultrasonography can be useful in guiding injections or barbotage, such as aspirating calcific deposits in the rotator cuff [20].
  • Ultrasonography is a low-cost alternative to MRI and arthrography for evaluating both skeletal and soft-tissue structures of the shoulder [27].
  • Ultrasonography can provide immediate, real-time visualization of the rotator cuff, biceps tendon, and calcific deposits [27].
  • Ultrasonography can be used to measure the subacromial space [27].
  • Ultrasonography can detect atrophy of rotator cuff muscles [27].
  • As a result of providing images in real-time, ultrasonography can evaluate impingement in various positions and motions [27].
  • Ultrasonography is highly operator dependent [27].
  • Ultrasonography is not as useful for evaluating labral tears or rotator cuff tears that are very small or larger than 3 cm [27].
  • The most commonly performed joint examination using ultrasonography is the shoulder examination [18].
  • Accuracy of rotator cuff ultrasonography depends on the skill of the scanner operator and an awareness of pitfalls encountered [18].
  • Ultrasonography can pick up on partial tears that may need treatment but have normal physical examination findings due to the ability to move the tendon with direct visualization [18].
  • In a study of 147 patients with rotator cuff calcific tendinitis treated with ultrasonography-guided lavage, 70% of shoulders resulted in significant reduction of symptoms [18].
  • Calcifications that were softer and middle-sized (12 to 17 mm) had more significant improvement after ultrasonography-guided lavage [18].
  • Better results after ultrasonography-guided lavage occurred in patients aged 30 to 40 years [18].
  • The sensitivity of ultrasonography for the detection of full-thickness rotator cuff tears is 98% [28].
  • The specificity of ultrasonography for the detection of full-thickness rotator cuff tears is 80% [28].
  • The positive predictive value of ultrasonography for the detection of full-thickness rotator cuff tears is 90% [28].
  • The negative predictive value of ultrasonography for the detection of full-thickness rotator cuff tears is 95% [28].
  • The accuracy of ultrasonography for the detection of full-thickness rotator cuff tears is 94% [28].
  • The sensitivity of MRI for the detection of full-thickness rotator cuff tears is 100% [28].
  • The specificity of MRI for the detection of full-thickness rotator cuff tears is 68% [28].
  • The positive predictive value of MRI for the detection of full-thickness rotator cuff tears is 85% [28].
  • The negative predictive value of MRI for the detection of full-thickness rotator cuff tears is 100% [28].
  • The accuracy of MRI for the detection of full-thickness rotator cuff tears is 89% [28].

General Imaging Principles

  • The shoulder is a three-dimensional structure that cannot be represented by a single planar view [23].
  • Critical relationships, such as the degree of centering of the humeral head, change with the position of the arm [23].
  • Shoulder pathology may be found in a large number of different bones and soft tissues [23].
  • Overlying and superimposed structures as well as metallic implants may complicate imaging the structures of interest [23].
  • Surgeons need to develop a judicious approach to imaging that yields necessary information while avoiding the tendency to "over-image" [23].
  • Proper radiographic technique is as important as proper surgical technique to achieve the desired outcome [2].
  • Radiograph technologists must know what is being sought in the images to ensure proper technique [2].
  • Many "axillary views" sent for consultation are taken without standardization, making it impossible to determine important features of the glenohumeral joint [2].

References

[2] Rockwood And Matsen S The Shoulder. Arthroscopic Management of Prearthritic and Arthritic Conditions of the Shoulder and the Postarthroplasty Shoulder > Radiographic Evaluation.

[3] Rockwood And Matsen S The Shoulder. Shoulder and Elbow Specialty Clinic Workers’ Survey > ANATOMY.

[4] A Lange Medical Book Current Diagnosis Treatment In Orthopedics Fifth Edition. 2Musculoskeletal Trauma Surgery > SHOULDER AND ARM INJURIES.

[5] Rockwood And Green S Fractures In Adults. 29: Principles of Nonunion and Bone Defect Treatment > Applied Anatomy Related to Scapular Fractures.

[6] Aaos Comprehensive Orthopaedic Review 3. Anatomy of the Shoulder, Arm, and Elbow > I. Shoulder.

[8] Rockwood And Matsen S The Shoulder. Risk Factors for Failure of Arthroscopic Stabilization > ANATOMY.

[10] Rockwood And Matsen S The Shoulder. Fractures, Dislocations, and Acquired Problems of the Shoulder in Children > FRACTURES OF THE PROXIMAL HUMERUS.

[11] Apley And Solomon S Concise System Of Orthopaedics And Trauma. DISORDERS OF THE ROTATOR CUFF.

[12] Aaos Comprehensive Orthopaedic Review 3. Fractures of the Clavicle, Scapula, and Glenoid > I. Clavicular Fractures.

[13] Rockwood And Matsen S The Shoulder. Developmental Anatomy of the Shoulder and Anatomy of the Glenohumeral Joint > Development of Individual Regions.

[15] Campbell S Operative Orthopaedics 4 Volume Set. RECONSTRUCTIVE PROCEDURES OF THE SHOULDER AND ELBOW IN ADULTS > ANATOMY AND BIOMECHANICS.

[16] Campbell S Operative Orthopaedics 4 Volume Set. ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH BONE-PATELLAR TENDON-BONE GRAFT > SHOULDER INJURIES > ANATOMY AND BIOMECHANICS.

[17] Miller S Review Of Orthopaedics. Genetics of musculoskeletal conditions and abnormalities are summarized in Table 1.27 > UPPER EXTREMITY > SHOULDER.

[18] Orthopaedic Knowledge Update Sports Medicine 6. Diagnostic Ultrasonography and Ultrasonography-­Guided Procedures > Annotated References.

[20] Apley And Solomon S Concise System Of Orthopaedics And Trauma. INVESTIGATION.

[22] A Lange Medical Book Current Diagnosis Treatment In Orthopedics Fifth Edition. 2Musculoskeletal Trauma Surgery > FRACTURES AND DISLOCATIONS AROUND THE SHOULDER.

[23] Rockwood And Matsen S The Shoulder. Developmental Anatomy of the Shoulder and Anatomy of the Glenohumeral Joint > SENIOR EDITOR COMMENTARY.

[25] Rockwood And Green S Fractures In Adults. 29: Principles of Nonunion and Bone Defect Treatment > Magnetic Resonance Imaging and Arthrography.

[27] Orthopaedic Knowledge Update 13 Ebook Without Multimedia. Shoulder Anatomy and Biomechanics, Clinical Evaluation, Imaging > Clinical Evaluation > Imaging.

[28] Aaos Comprehensive Orthopaedic Review 3. Imaging of the Shoulder and Elbow > I. Shoulder.

Creative Commons BY-NC 4.0

CC Creative Commons licence
BY Attribution — you must credit the source
NC NonCommercial — not for commercial use

Attribution-NonCommercial 4.0 International


Creative Commons Corporation ("Creative Commons") is not a law firm and does not provide legal services or legal advice. Distribution of Creative Commons public licenses does not create a lawyer-client or other relationship. Creative Commons makes its licenses and related information available on an "as-is" basis. Creative Commons gives no warranties regarding its licenses, any material licensed under their terms and conditions, or any related information. Creative Commons disclaims all liability for damages resulting from their use to the fullest extent possible.

Using Creative Commons Public Licenses

Creative Commons public licenses provide a standard set of terms and conditions that creators and other rights holders may use to share original works of authorship and other material subject to copyright and certain other rights specified in the public license below. The following considerations are for informational purposes only, are not exhaustive, and do not form part of our licenses.

Considerations for licensors: Our public licenses are intended for use by those authorized to give the public permission to use material in ways otherwise restricted by copyright and certain other rights. Our licenses are irrevocable. Licensors should read and understand the terms and conditions of the license they choose before applying it. Licensors should also secure all rights necessary before applying our licenses so that the public can reuse the material as expected. Licensors should clearly mark any material not subject to the license. This includes other CC- licensed material, or material used under an exception or limitation to copyright. More considerations for licensors: wiki.creativecommons.org/Considerations_for_licensors

Considerations for the public: By using one of our public licenses, a licensor grants the public permission to use the licensed material under specified terms and conditions. If the licensor's permission is not necessary for any reason--for example, because of any applicable exception or limitation to copyright--then that use is not regulated by the license. Our licenses grant only permissions under copyright and certain other rights that a licensor has authority to grant. Use of the licensed material may still be restricted for other reasons, including because others have copyright or other rights in the material. A licensor may make special requests, such as asking that all changes be marked or described. Although not required by our licenses, you are encouraged to respect those requests where reasonable. More considerations for the public: wiki.creativecommons.org/Considerations_for_licensees


Creative Commons Attribution-NonCommercial 4.0 International Public License

By exercising the Licensed Rights (defined below), You accept and agree to be bound by the terms and conditions of this Creative Commons Attribution-NonCommercial 4.0 International Public License ("Public License"). To the extent this Public License may be interpreted as a contract, You are granted the Licensed Rights in consideration of Your acceptance of these terms and conditions, and the Licensor grants You such rights in consideration of benefits the Licensor receives from making the Licensed Material available under these terms and conditions.

Section 1 -- Definitions.

a. Adapted Material means material subject to Copyright and Similar Rights that is derived from or based upon the Licensed Material and in which the Licensed Material is translated, altered, arranged, transformed, or otherwise modified in a manner requiring permission under the Copyright and Similar Rights held by the Licensor. For purposes of this Public License, where the Licensed Material is a musical work, performance, or sound recording, Adapted Material is always produced where the Licensed Material is synched in timed relation with a moving image.

b. Adapter's License means the license You apply to Your Copyright and Similar Rights in Your contributions to Adapted Material in accordance with the terms and conditions of this Public License.

c. Copyright and Similar Rights means copyright and/or similar rights closely related to copyright including, without limitation, performance, broadcast, sound recording, and Sui Generis Database Rights, without regard to how the rights are labeled or categorized. For purposes of this Public License, the rights specified in Section 2(b)(1)-(2) are not Copyright and Similar Rights.

d. Effective Technological Measures means those measures that, in the absence of proper authority, may not be circumvented under laws fulfilling obligations under Article 11 of the WIPO Copyright Treaty adopted on December 20, 1996, and/or similar international agreements.

e. Exceptions and Limitations means fair use, fair dealing, and/or any other exception or limitation to Copyright and Similar Rights that applies to Your use of the Licensed Material.

f. Licensed Material means the artistic or literary work, database, or other material to which the Licensor applied this Public License.

g. Licensed Rights means the rights granted to You subject to the terms and conditions of this Public License, which are limited to all Copyright and Similar Rights that apply to Your use of the Licensed Material and that the Licensor has authority to license.

h. Licensor means the individual(s) or entity(ies) granting rights under this Public License.

i. NonCommercial means not primarily intended for or directed towards commercial advantage or monetary compensation. For purposes of this Public License, the exchange of the Licensed Material for other material subject to Copyright and Similar Rights by digital file-sharing or similar means is NonCommercial provided there is no payment of monetary compensation in connection with the exchange.

j. Share means to provide material to the public by any means or process that requires permission under the Licensed Rights, such as reproduction, public display, public performance, distribution, dissemination, communication, or importation, and to make material available to the public including in ways that members of the public may access the material from a place and at a time individually chosen by them.

k. Sui Generis Database Rights means rights other than copyright resulting from Directive 96/9/EC of the European Parliament and of the Council of 11 March 1996 on the legal protection of databases, as amended and/or succeeded, as well as other essentially equivalent rights anywhere in the world.

l. You means the individual or entity exercising the Licensed Rights under this Public License. Your has a corresponding meaning.

Section 2 -- Scope.

a. License grant.

1. Subject to the terms and conditions of this Public License, the Licensor hereby grants You a worldwide, royalty-free, non-sublicensable, non-exclusive, irrevocable license to exercise the Licensed Rights in the Licensed Material to:

a. reproduce and Share the Licensed Material, in whole or in part, for NonCommercial purposes only; and

b. produce, reproduce, and Share Adapted Material for NonCommercial purposes only.

2. Exceptions and Limitations. For the avoidance of doubt, where Exceptions and Limitations apply to Your use, this Public License does not apply, and You do not need to comply with its terms and conditions.

3. Term. The term of this Public License is specified in Section 6(a).

4. Media and formats; technical modifications allowed. The Licensor authorizes You to exercise the Licensed Rights in all media and formats whether now known or hereafter created, and to make technical modifications necessary to do so. The Licensor waives and/or agrees not to assert any right or authority to forbid You from making technical modifications necessary to exercise the Licensed Rights, including technical modifications necessary to circumvent Effective Technological Measures. For purposes of this Public License, simply making modifications authorized by this Section 2(a) (4) never produces Adapted Material.

5. Downstream recipients.

a. Offer from the Licensor -- Licensed Material. Every recipient of the Licensed Material automatically receives an offer from the Licensor to exercise the Licensed Rights under the terms and conditions of this Public License.

b. No downstream restrictions. You may not offer or impose any additional or different terms or conditions on, or apply any Effective Technological Measures to, the Licensed Material if doing so restricts exercise of the Licensed Rights by any recipient of the Licensed Material.

6. No endorsement. Nothing in this Public License constitutes or may be construed as permission to assert or imply that You are, or that Your use of the Licensed Material is, connected with, or sponsored, endorsed, or granted official status by, the Licensor or others designated to receive attribution as provided in Section 3(a)(1)(A)(i).

b. Other rights.

1. Moral rights, such as the right of integrity, are not licensed under this Public License, nor are publicity, privacy, and/or other similar personality rights; however, to the extent possible, the Licensor waives and/or agrees not to assert any such rights held by the Licensor to the limited extent necessary to allow You to exercise the Licensed Rights, but not otherwise.

2. Patent and trademark rights are not licensed under this Public License.

3. To the extent possible, the Licensor waives any right to collect royalties from You for the exercise of the Licensed Rights, whether directly or through a collecting society under any voluntary or waivable statutory or compulsory licensing scheme. In all other cases the Licensor expressly reserves any right to collect such royalties, including when the Licensed Material is used other than for NonCommercial purposes.

Section 3 -- License Conditions.

Your exercise of the Licensed Rights is expressly made subject to the following conditions.

a. Attribution.

1. If You Share the Licensed Material (including in modified form), You must:

a. retain the following if it is supplied by the Licensor with the Licensed Material:

i. identification of the creator(s) of the Licensed Material and any others designated to receive attribution, in any reasonable manner requested by the Licensor (including by pseudonym if designated);

ii. a copyright notice;

iii. a notice that refers to this Public License;

iv. a notice that refers to the disclaimer of warranties;

v. a URI or hyperlink to the Licensed Material to the extent reasonably practicable;

b. indicate if You modified the Licensed Material and retain an indication of any previous modifications; and

c. indicate the Licensed Material is licensed under this Public License, and include the text of, or the URI or hyperlink to, this Public License.

2. You may satisfy the conditions in Section 3(a)(1) in any reasonable manner based on the medium, means, and context in which You Share the Licensed Material. For example, it may be reasonable to satisfy the conditions by providing a URI or hyperlink to a resource that includes the required information.

3. If requested by the Licensor, You must remove any of the information required by Section 3(a)(1)(A) to the extent reasonably practicable.

4. If You Share Adapted Material You produce, the Adapter's License You apply must not prevent recipients of the Adapted Material from complying with this Public License.

Section 4 -- Sui Generis Database Rights.

Where the Licensed Rights include Sui Generis Database Rights that apply to Your use of the Licensed Material:

a. for the avoidance of doubt, Section 2(a)(1) grants You the right to extract, reuse, reproduce, and Share all or a substantial portion of the contents of the database for NonCommercial purposes only;

b. if You include all or a substantial portion of the database contents in a database in which You have Sui Generis Database Rights, then the database in which You have Sui Generis Database Rights (but not its individual contents) is Adapted Material; and

c. You must comply with the conditions in Section 3(a) if You Share all or a substantial portion of the contents of the database.

For the avoidance of doubt, this Section 4 supplements and does not replace Your obligations under this Public License where the Licensed Rights include other Copyright and Similar Rights.

Section 5 -- Disclaimer of Warranties and Limitation of Liability.

a. UNLESS OTHERWISE SEPARATELY UNDERTAKEN BY THE LICENSOR, TO THE EXTENT POSSIBLE, THE LICENSOR OFFERS THE LICENSED MATERIAL AS-IS AND AS-AVAILABLE, AND MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY KIND CONCERNING THE LICENSED MATERIAL, WHETHER EXPRESS, IMPLIED, STATUTORY, OR OTHER. THIS INCLUDES, WITHOUT LIMITATION, WARRANTIES OF TITLE, MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, NON-INFRINGEMENT, ABSENCE OF LATENT OR OTHER DEFECTS, ACCURACY, OR THE PRESENCE OR ABSENCE OF ERRORS, WHETHER OR NOT KNOWN OR DISCOVERABLE. WHERE DISCLAIMERS OF WARRANTIES ARE NOT ALLOWED IN FULL OR IN PART, THIS DISCLAIMER MAY NOT APPLY TO YOU.

b. TO THE EXTENT POSSIBLE, IN NO EVENT WILL THE LICENSOR BE LIABLE TO YOU ON ANY LEGAL THEORY (INCLUDING, WITHOUT LIMITATION, NEGLIGENCE) OR OTHERWISE FOR ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, CONSEQUENTIAL, PUNITIVE, EXEMPLARY, OR OTHER LOSSES, COSTS, EXPENSES, OR DAMAGES ARISING OUT OF THIS PUBLIC LICENSE OR USE OF THE LICENSED MATERIAL, EVEN IF THE LICENSOR HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH LOSSES, COSTS, EXPENSES, OR DAMAGES. WHERE A LIMITATION OF LIABILITY IS NOT ALLOWED IN FULL OR IN PART, THIS LIMITATION MAY NOT APPLY TO YOU.

c. The disclaimer of warranties and limitation of liability provided above shall be interpreted in a manner that, to the extent possible, most closely approximates an absolute disclaimer and waiver of all liability.

Section 6 -- Term and Termination.

a. This Public License applies for the term of the Copyright and Similar Rights licensed here. However, if You fail to comply with this Public License, then Your rights under this Public License terminate automatically.

b. Where Your right to use the Licensed Material has terminated under Section 6(a), it reinstates:

1. automatically as of the date the violation is cured, provided it is cured within 30 days of Your discovery of the violation; or

2. upon express reinstatement by the Licensor.

For the avoidance of doubt, this Section 6(b) does not affect any right the Licensor may have to seek remedies for Your violations of this Public License.

c. For the avoidance of doubt, the Licensor may also offer the Licensed Material under separate terms or conditions or stop distributing the Licensed Material at any time; however, doing so will not terminate this Public License.

d. Sections 1, 5, 6, 7, and 8 survive termination of this Public License.

Section 7 -- Other Terms and Conditions.

a. The Licensor shall not be bound by any additional or different terms or conditions communicated by You unless expressly agreed.

b. Any arrangements, understandings, or agreements regarding the Licensed Material not stated herein are separate from and independent of the terms and conditions of this Public License.

Section 8 -- Interpretation.

a. For the avoidance of doubt, this Public License does not, and shall not be interpreted to, reduce, limit, restrict, or impose conditions on any use of the Licensed Material that could lawfully be made without permission under this Public License.

b. To the extent possible, if any provision of this Public License is deemed unenforceable, it shall be automatically reformed to the minimum extent necessary to make it enforceable. If the provision cannot be reformed, it shall be severed from this Public License without affecting the enforceability of the remaining terms and conditions.

c. No term or condition of this Public License will be waived and no failure to comply consented to unless expressly agreed to by the Licensor.

d. Nothing in this Public License constitutes or may be interpreted as a limitation upon, or waiver of, any privileges and immunities that apply to the Licensor or You, including from the legal processes of any jurisdiction or authority.


Creative Commons is not a party to its public licenses. Notwithstanding, Creative Commons may elect to apply one of its public licenses to material it publishes and in those instances will be considered the “Licensor.” The text of the Creative Commons public licenses is dedicated to the public domain under the CC0 Public Domain Dedication. Except for the limited purpose of indicating that material is shared under a Creative Commons public license or as otherwise permitted by the Creative Commons policies published at creativecommons.org/policies, Creative Commons does not authorize the use of the trademark "Creative Commons" or any other trademark or logo of Creative Commons without its prior written consent including, without limitation, in connection with any unauthorized modifications to any of its public licenses or any other arrangements, understandings, or agreements concerning use of licensed material. For the avoidance of doubt, this paragraph does not form part of the public licenses.

Creative Commons may be contacted at creativecommons.org.